Hydrogen peroxide sterile delivery window
TL-STB1000 Hydrogen Peroxide Aseptic Transfer Window
TL-STB1000 Hydrogen Peroxide Aseptic Transfer Window
Large capacity aseptic transfer solution, using hydrogen peroxide vapor low-temperature sterilization technology, meeting high standard clean transfer requirements
Product Features
- 1 TL-STB1000 aseptic transfer chamber uses hydrogen peroxide vapor as a biological decontaminant, a decontamination process at low temperature and atmospheric pressure, efficient, environmentally friendly, the hydrogen peroxide gas sterilization method is also included in the "Sterilization Methods" in the pharmacopoeias of the United States, China, and other countries;
- 2 Automatically controlled by Siemens Programmable Logic Controller (PLC) for each stage of equipment operation, ensuring stable and professional system operation;
- 3 During biological decontamination and exhaust stages, air sent into the chamber is filtered through H14 class HEPA filters to prevent material from contamination;
- 4 Equipped with an independent ventilation and residual removal unit, including fan, pneumatic butterfly valve, and exhaust pipeline, capable of quickly replacing hydrogen peroxide gas in the chamber and preventing it from entering the HVAC system;
- 5 Feeding and discharging double-leaf door structure, with pneumatic sealing, pneumatic locking, and dual-door interlock function during operation;
- 6 Internal trolley-type material rack and track system, facilitating material loading and unloading;
- 7 The system has real-time monitoring functions for temperature, humidity, and pressure inside the chamber, optional hydrogen peroxide gas concentration, airborne particles, and viable particle detection ports;
- 8 Easy to validate, with complete IQ, OQ, PQ documents, and can provide on-site validation;
- 9 Biological decontamination cycle parameter data saving and printing function;
Core Advantage: TL-STB1000 adopts advanced Siemens PLC automatic control system and H14 class HEPA filtration technology, ensuring high professionalism and verifiability of the aseptic transfer process.
Technical Specifications
| Parameter Item | Specification |
|---|---|
| Chamber Material | SUS 304 Stainless Steel / SUS 316 Stainless Steel |
| HEPA Filter Level | Supply Air HEPA: H14 Class / Exhaust Air HEPA: H14 Class |
| Alarm Functions | High/Low Pressure Alarm; Door Open Timeout Alarm; Door Lock Alarm |
| Biological Decontamination Cycle Time | < 90min |
| Biological Decontamination Efficacy | 1g (Geobacillus stearothermophilus, ATCC12980 or ATCC7953) |
| VHPS Residual Level | ≤ 1ppm |
| Chamber Cleanliness (Particle Level) | Class A |
| Hydrogen Peroxide Gas Decontamination Cycle Interface | DN32 Sanitary Interface |
| Exhaust Port Interface | DN50 Sanitary Interface |
| Transfer Door Diameter | 28cm |
Equipment Dimension Specifications
Note: The above technical specifications are standard configurations, can be customized and adjusted according to specific customer requirements.