Hard cabin sterile isolator

Aseptic sampling isolator (open type)

Word:[Big][Middle][Small] QR Code 2026-1-9     Viewed:    
无菌取样隔离器(开放式)
Aseptic Sampling Isolator (Open Type) | Product Details

Basic Introduction

The aseptic sampling isolation system (open type) consists of a Vaporized Hydrogen Peroxide (VHP) aseptic transfer window and an ORABS system. The VHP aseptic transfer window integrates sterilization and self-cleaning functions, meeting the professional equipment requirements for clean/aseptic material transfer between different clean areas, complying with relevant regulations such as GMP.

The ORABS system consists of three parts: feeding window, operation window, and discharging window. The feeding window is mainly used for material temporary storage and opening; the operation window is used for weighing and sampling; the discharging window is used for crimping.

Our company can customize isolation systems and RABS systems according to customer requirements.

Performance Features

  • 1 The VHP aseptic transfer window mainly consists of window body, VHP sterilization system, double-leaf door, HEPA filter, centrifugal fan, temperature and humidity sensors, pressure differential gauge, needle printer, micro differential pressure sensor, alarm device, interlock device, and control components.
  • 2 The ORABS system mainly consists of window body, HEPA filter, centrifugal fan, online wind speed sensor, pressure differential gauge, electronic scale, etc. The window is enclosed with transparent acrylic panels on both sides, with glove operation ports, convenient for production operations, reducing the possibility of environmental contamination inside the ORABS by opening doors during operation.
  • 3 The ORABS system should be placed in a Class B area.

Technical Specifications

Power Supply
AC220±22V/50±1Hz
Power
4000W
Sterilizer Vaporization Temperature
≤90℃
Noise
≤70dB(A)
Sterilant
30% Hydrogen Peroxide Solution
Kill Rate
Killing rate of 6-log for Geobacillus stearothermophilus and Bacillus subtilis
Transfer Window Volume
0.80 cubic meters
Air Velocity
0.36~0.54m/s
HEPA Filter
H14 Class

Equipment Dimension Parameters

Component Name External Dimensions (mm) Internal Dimensions (mm)
Transfer Window 1100×1100×2000 800×800×1000
Feeding Window 1890×1100×2200 -
Operation Window 1280×690×2200 -
Discharging Window 1280×690×2200 -

Note: The above technical specifications are standard configurations, can be customized and adjusted according to specific customer requirements.

Application Areas

1 Pharmaceutical Industry: Sampling of sterile bulk drugs.
2 Customization for other areas with requirements.

Customization Service: Can provide customized aseptic sampling isolation system solutions according to special requirements of different industries.

System Components

VHP Aseptic Transfer Window
Vaporized Hydrogen Peroxide Sterilization System
Feeding Window
Material Temporary Storage and Opening
Operation Window
Weighing and Sampling Operations
Discharging Window
Crimping and Material Output
Control System
Automated Control and Monitoring